Experience | Guidelines | News | Products | Other
This is a PatientPlus article. PatientPlus articles are written for doctors and so the language can be technical. However, some people find that they add depth to the articles found in the other sections of this website which are written for non-medical people.
Intrapartum Fetal Monitoring
Post your experienceThere are a wide range of clinical techniques and investigations available to healthcare professionals to survey the condition of a fetus in the womb. These may be deployed from early pregnancy to birth. They range from the use of traditional equipment such as a Pinard stethoscope that allows direct auscultation of the fetal heart beat, to ultrasound imaging of the fetus that gives an ever-increasing amount of morphological and functional data about the unborn child.
This article deals with those monitoring techniques that are used immediately preceding or during childbirth – known as intrapartum fetal monitoring. Current techniques rely predominantly on the use of electronic fetal monitoring through the use of cardiotocography (CTG). This technique allows detection and analysis of the fetal heart rate (via Doppler ultrasound or direct fetal ECG measurement with a fetal scalp electrode) and a semi-quantitative analysis of myometrial activity and contractions. The interpretation of the data collected depends on the relationship between the two traces.
The technique became widely used from the 1960s onwards. Prior to this there was very little that could be discerned about the state of the fetus. The new technology was widely embraced as an undoubted 'good thing' that would lead to better outcomes and reduce the incidence of conditions such as cerebral palsy that were postulated to be largely due to episodes of intrapartum cerebral ischaemia. Unfortunately, subsequent evidence has not borne out this optimism.
It seems unlikely that more than a small minority of cases of cerebral palsy are directly attributable to avoidable intrapartum fetal hypoxia:1,2
- Low gestational age (24-30 weeks) postnatal dexamethasone use, patent ductus arteriosus, severe hyaline membrane disease, resuscitation in the delivery room and intraventricular haemorrhage have all been shown to be associated with higher rates of cerebral palsy, whereas antenatal corticosteroid is associated with a lower rate.3
- There appears to be at best a tenuous connection between cardiotocographic findings, what they signify about the fetal condition and any improvement in outcomes as a result of intervention based upon them.4
- A Cochrane systematic review has found no evidence of the usefulness of cardiotocography in antepartum fetal assessment, and even a trend to an increase in perinatal deaths in patients being managed using cardiotocography (odds ratio 2.85, 95% confidence interval 0.99 to 7.12).5
- The widespread use of electronic fetal monitoring as part of the management of normal labour has been discarded, as a result of professional and lay voices alike.6,7
- A Cochrane meta-analysis showed that the use of routine CTG had a minor beneficial effect on the incidence of neonatal seizures, but increased instrumental and caesarean deliveries significantly, calling into question its continuing routine use, against current guideline advice.8
- Poor understanding of the patho-physiological background of the measurements collected
- Indirect fetal monitoring treated by some as a direct monitor of fetal condition
- Numerous technical problems that affect its usefulness
- Variability in measurement and recording techniques
- Qualitative nature of information obtained, requiring complex pattern recognition
- An absence of agreed systems of classification
- Confusion about the many influences on the fetal heart rhythm and rate
- Significant intra- and inter-observer variation in interpretation of data
- Low validity of findings and high false-positive finding rate
- A screening investigation to detect the presence of fetal distress being misused as a diagnostic tool
- Inevitably leads to an increase in instrumental or surgical deliveries, due to the high false-positive rate
- No satisfactory criteria on how, when and who to monitor
- Contributes significantly to medico-legal vulnerability of doctors who manage childbirth
It can be used as a fairly reliable screening test to indicate a normal condition of the fetus. If the fetal heart rate and pattern is normal it is virtually certain that the fetus is not suffering hypoxia, acidaemia or other causes of distress. It is this useful aspect of cardiotocography that led to its widespread adoption. Unfortunately, its ubiquitous presence on labour wards has led to its overzealous use and overcomplicated interpretation, without a solid evidence-based grounding. The table below lists features of the CTG that would lead one to consider it reassuring, non-reassuring or abnormal.
CTG fetal heart rate features and classification7 |
||||
|---|---|---|---|---|
| Baseline HR/bpm | HR Variability/bpm | Decelerations in fetal HR | Accelerations in fetal heart rate | |
| Reassuring | 110–160 | ≥5 | Absent | Present |
| Non-reassuring | 100–109 or 161–180 | <5 for ≥40 mins but <90 mins | Early decelerations or variable decelerations or single prolonged deceleration lasting ≤3 mins | Absence of accelerations in an otherwise normal CTG is of uncertain significance |
| Abnormal | <100 or >180 or sinusoidal pattern for ≥10 mins | <5 for ≥90 mins | Atypical variable decelerations or late decelerations or single prolonged deceleration lasting >3 mins | Absence of accelerations in an otherwise normal CTG is of uncertain significance |
|
- If repeated accelerations are present with reduced variability, the FHR trace should be regarded as reassuring.
- True early uniform decelerations are rare and benign, and therefore they are not significant.
- Most decelerations in labour are variable.
- If a bradycardia occurs in the baby for more than 3 minutes, urgent medical aid should be sought and preparations should be made to urgently expedite the birth of the baby. This could include moving the woman to theatre if the fetal heart has not recovered by 9 minutes. If the fetal heart recovers within 9 minutes the decision to deliver should be reconsidered, in conjunction with the woman if reasonable.
- A tachycardia in the baby of 160–180 bpm, where accelerations are present and no other adverse features appear, should not be regarded as suspicious. However, an increase in the baseline heart rate, even within the normal range, with other non-reassuring or abnormal features should increase concern.
The following table of antenatal and intrapartum risk factors should prompt the use of continuous electronic fetal monitoring, according to algorithmic guidelines issued by NICE & RCOG.7 The lists are not exhaustive and other risk factors may prompt the use of continuous monitoring.
| All other patients having normal labour without associated risk factors should not have continuous electronic fetal monitoring. |
All decisions to use continuous electronic fetal monitoring should be discussed with the patient and the reasons for offering it should be outlined. It is important to note that these criteria are only for the offering of continuous electronic fetal monitoring, not its mandatory use, and the pregnant mother is entitled to have the last word on whether or not she wishes to use it. Where it has been recommended and declined, this fact should be carefully documented in the partogram and clinical record.
Indications for continuous electronic fetal monitoring during pregnancy |
||
|---|---|---|
| Antenatal maternal risk factors | Antenatal fetal risk factors | Intrapartum risk factors |
| Previous caesarean section | Fetal growth restriction | Augmentation of labour using oxytocin |
| Pre-eclampsia | Prematurity | Epidural analgesia |
| Post-term pregnancy (>42 weeks) | Oligohydramnios | Vaginal bleeding during labour |
| Prolonged membrane rupture (>24 hours) | Abnormal placental blood flow as indicated by Doppler artery velocitometry | Maternal pyrexia |
| Induced labour | Multiple pregnancies | Fresh meconium-stained liquor |
| Diabetes | Meconium-stained liquor | |
| Ante-partum haemorrhage | Breech presentation | |
| Other significant maternal medical disease | ||
A Cochrane systematic review has indicated that when continuous electronic fetal monitoring is indicated, the use of a fetal scalp electrode to measure fetal heart rate and allow ST waveform analysis by ECG, along with CTG, results in fewer blood samples taken from the baby's scalp, less surgical assistance and better oxygen levels at birth than CTG alone.10
For a suspicious CTG:7
|
For a pathological CTG:7
|
The table below gives the current RCOG & NICE guidance on appropriate responses to fetal scalp pH results.7,7
Fetal scalp pH results and apropriate courses of action |
|
|---|---|
| Fetal blood sample (FBS) result/pH | Subsequent action |
| ≥7.25 | Repeat FBS if fetal HR abnormalities persist |
| 7.21–7.24 | Repeat FBS within 30 minutes or consider delivery if rapid fall in pH since last sample |
| ≤7.20 | Delivery indicated |
| All fetal scalp blood pH estimations should be interpreted taking into account the previous pH measurement, the rate of progress in labour and the clinical features of the mother and baby. |
|
- A Swedish RCT examined the effectiveness of pH analysis of fetal scalp blood compared with lactate analysis in identifying hypoxia in labour to prevent acidaemia at birth. Most outcome measures showed NO statistical difference e.g. acidaemia rate at birth, APGAR score, but the large difference was in the rate of failure of obtaining a sample - much less with lactate (smaller sample required) than pH.13
- Vibroacoustic stimulation of the fetus as an adjunct to CTG may improve its utility by reducing the incidence of non-reactivity (if, for example, the fetus is sleeping).14 However, a recent Cochrane systematic review found that there is insufficient evidence from randomised trials on which to base recommendations for use of vibroacoustic stimulation in the evaluation of fetal well-being in labour in the presence of a non-reassuring cardiotocographic trace.15
- A new method of fetal pulse oximetry may be useful for reducing caesarean section rates in the presence of non-reassuring CTG findings.16
- The use of automated computerised analysis systems to interpret CTGs and avoid intra- and inter-observer variability is being shown to have some beneficial effects in trials and offers hope of more consistent criteria for intervention on the basis of CTG findings in future.17
- The combination of auto-analysis of fetal pulse oximetry and CTG data is also being investigated with some evidence of usefulness.18
Continuous electronic fetal monitoring is a useful intrapartum tool in experienced hands, if used selectively and according to evidence-based guidelines. It should not be used routinely as this is one of the factors that has pushed up the instrumental and caesarean delivery rates in the developed world. Its use should continue to be investigated by carefully designed randomised controlled trials to optimise its utility and help improve and develop guidelines for its use.
Document references
- Keogh JM, Badawi N; The origins of cerebral palsy. Curr Opin Neurol. 2006 Apr;19(2):129-34. [abstract]
- Blair E, Watson L; Epidemiology of cerebral palsy. Semin Fetal Neonatal Med. 2006 Apr;11(2):117-25. Epub 2005 Dec 9. [abstract]
- Vincer MJ, Allen AC, Joseph KS, et al; Increasing prevalence of cerebral palsy among very preterm infants: a population-based study. Pediatrics. 2006 Dec;118(6):e1621-6. Epub 2006 Oct 30. [abstract]
- Parer JT, King T, Flanders S, et al; Fetal acidemia and electronic fetal heart rate patterns: Is there evidence of an association? J Matern Fetal Neonatal Med. 2006 May;19(5):289-94. [abstract]
- Pattison N, McCowan L; Cardiotocography for antepartum fetal assessment. Cochrane Database Syst Rev. 2000;(2):CD001068. [abstract]
- Kitzinger S, Green JM, Chalmers B, et al; Why do women go along with this stuff? Birth. 2006 Jun;33(2):154-8. [abstract]
- Intrapartum care, NICE Clinical Guideline (2007)
- Thacker SB, Stroup D, Chang M; Continuous electronic heart rate monitoring for fetal assessment during labor. Cochrane Database Syst Rev. 2001;(2):CD000063. [abstract]
- van Geijn H; Cardiotocography, ObGyn.net, 1998
; An objective overview of the use of cardiotocography from a practising obstetrician - Neilson JP; Fetal electrocardiogram (ECG) for fetal monitoring during labour. Cochrane Database Syst Rev. 2003;(2):CD000116. [abstract]
- Thomas J, Paranjothy S, James D; National cross sectional survey to determine whether the decision to delivery interval is critical in emergency caesarean section. BMJ. 2004 Mar 20;328(7441):665. Epub 2004 Mar 15. [abstract]
- MacKenzie IZ, Cooke I; What is a reasonable time from decision-to-delivery by caesarean section? Evidence from 415 deliveries. BJOG. 2002 May;109(5):498-504. [abstract]
- Wiberg-Itzel E, Lipponer C, Norman M, et al; Determination of pH or lactate in fetal scalp blood in management of intrapartum fetal distress: randomised controlled multicentre trial. BMJ. 2008 Jun 7;336(7656):1284-7. Epub 2008 May 25. [abstract]
- Tan KH, Smyth R; Fetal vibroacoustic stimulation for facilitation of tests of fetal wellbeing. Cochrane Database Syst Rev. 2001;(1):CD002963. [abstract]
- East CE, Smyth R, Leader LR, et al; Vibroacoustic stimulation for fetal assessment in labour in the presence of a nonreassuring fetal heart rate trace. Cochrane Database Syst Rev. 2005 Apr 18;(2):CD004664. [abstract]
- East CE, Chan FY, Colditz PB; Fetal pulse oximetry for fetal assessment in labour. Cochrane Database Syst Rev. 2004 Oct 18;(4):CD004075. [abstract]
- Ojala K, Vaarasmaki M, Makikallio K, et al; A comparison of intrapartum automated fetal electrocardiography and conventional cardiotocography--a randomised controlled study. BJOG. 2006 Apr;113(4):419-23. [abstract]
- Salamalekis E, Siristatidis C, Vasios G, et al; Fetal pulse oximetry and wavelet analysis of the fetal heart rate in the evaluation of abnormal cardiotocography tracings. J Obstet Gynaecol Res. 2006 Apr;32(2):135-9. [abstract]
Internet and further reading
- National Childbirth Trust
- Zeldin AS, Ratanawongsa B, Bassano ATF; Cerebral Palsy. eMedicine, March 2007.
Document ID: 1063
Document Version: 22
Document Reference: bgp220
Last Updated: 28 Jun 2008
Planned Review: 28 Jun 2010
The authors and editors of this article are employed to create accurate and up to date content reflecting reliable research evidence, guidance and best clinical practice. They are free from any commercial conflicts of interest. Find out more about updating.
Patient UK Hearing Impairment Survey
Patient UK are grateful to the 550 people who took part in this survey.
To see the results click here.
If you'd like to leave your feedback, please go to our interactive forum.
Experience | Guidelines | News | Products | Other
Related pages in Patient UK
Your Experience (^ top of page)
Please add your experience about this condition / medicineUK guidelines related to this topic (^ top of page)
Guidelines on Fetal MonitoringPatient UK Newspaper (^ top of page)
Recent related news items
Foetal heart rate monitor warning
Foetal kick charts 'inaccurate'
Cold 'cuts baby brain damage'
Cooling 'cuts baby brain damage'All news by related topic
Fetal Monitoring newsRelated Products (^ top of page)
Medical equipment

Books

Other - Useful resources (^ top of page)
Pictures, diagrams, photos, images, etc.Evidence based medicine
Online textbooks and journals
UK Guidelines
Online Videos
Medline
Other good health sites
Want to search some more? Use the Google Search box below to search our site.
Disclaimer: Patient UK has no control over the content of any external links above. Inclusion does not imply endorsement by Patient UK.
Want to advertise on this site? Find out how >>
Here you can follow a link to view existing patient experiences on this subject, or to add your own
This will offer you the usual PDF options i.e. document navigation, search, zoom and formatted print
Note: this is the best way to print the document
Note: this will open in a new window
Note: this will open in a new window
Here you can follow a link to view existing patient experiences on this subject, or to add your own
This will offer you the usual PDF options i.e. document navigation, search, zoom and formatted print
Note: this is the best way to print the document
Note: this will open in a new window
Note: this will open in a new window




